spn-58[1]

SPN 58

Share on:

Powerful features

  • Body weight mode input
  • Automatically calibrate syringe
  • Automatically identify syringe size
  • 3-level occlusion alarm limits
  • Automatically reduce impact dose release after occlusion
  • Delivered volume inquiry
  • KVO flow rate (0.1-1.0mL/h Adjustable)
  • Multi-direction clamp for either horizontal or vertical fixation
  • Operation error prompt
  • Fast rate control
  • “Purge” button activate safe protection
  • Flow rate setting button lock
  • Abundant optic and sound alarm
  • AC and internal power indication
  • Standard RS232 interface
  • 1000 history records


Brand: .

Product Description

Technical data SPN 58

STERNMED SYRINGE PUMP(Single-Channel)

Syringes

10, 20, 30, 50 mL of calibrated syringes

Flow rate

 

0.1~1500 mL/h (specified value 0.1 999.9 mL/h, 0.1 mL/h step)          50 mL syringe
(specified value 1000 mL/h or above, 1mL/h step)
0.1~900 mL/h (0.1 mL/h step)                                                                     30 mL syringe
0.1~600 mL/h (0.1 mL/h step)                                                                     20 mL syringe
0.1~400 mL/h (0.1 mL/h step)                                                                     10 mL syringe

Purge rate

1500 mL/h                                                                                                        50mL syringe
900 mL/h                                                                                                          30mL syringe
600 mL/h                                                                                                          20mL syringe
400 m/L h                                                                                                         10mL syringe

Accuracy of flow rate
(Nominal value)

Accuracy including syringe: within +2%
Mechanical accuracy: within +1%

Display range of volume delivered

0.1~9999 mL  (0.1~999.9 mL, 0.1 mL step)  (1000~9999 mL, 1mL step)

Setting of delivery limit

0.1~9999 mL  (0.1~999.9 mL, 0.1 mL step)  (1000~9999 mL, 1mL step)


Occlusion detection
pressure

‘H’: 800+200 mmHg ( 106.7+26.7 kPa) or 1.09+0.27 kgf/cm2
‘C’: 500+100 mmHg ( 66.7+13.3 kPa) or 0.68+0.13 kgf/cm2
‘L’: 300+100 mmHg (40.7+13.3 kPa) or 0.41+00.13 kgf/cm2

 

Alarms

‘Occlusion’, ‘Nearly empty’, Finish’, ‘Syringe dislocated’
‘Plunger/Clutch disengaged’, ‘Battery low’, ‘Battery exhausted’,
Power cable disconnected’, Flow rate over limit’, Volume over limit’, System error’.
‘Non-operation time over’

 

Power 

AC100+200V  10%, 50~60 Hz + 1 Hz
Inner NiMH battery: DC12V
Continuous operation time of battery :over 4 hours for dual channel when battery is recharged for 12 hours, at flow rate of 5 mL/h.

Operation conditions

Ambient temperature: 5 to 40 °C,
Relative humidity: 20% to 80%

Storage conditions

Ambient temperature: -20 to 55 °C,
Relative Humidity: <95%

Electric shock protection/class

Class and internal power supply: Type CF, Continuous operation

Water proof level

IPX4

Weight

2.25kg (with clamp base)

Dimension

306mm (Width) x135mm (Height) x 127mm (Depth)

Syringes

10, 20, 30, 50 mL of calibrated syringes