Technical Specification MACS 20 |
Sternmed Patient Monitor |
|
ECG Measurement |
|
|
Frequency response |
Monitoring mode: 0.5~40Hz (+0.4dB) ( -3.0dB) Diagnosis mode: 0.05~75Hz (+0.4dB) ( -3.0dB) |
|
CMRR |
≥90dB(diagnostic mode) ≥105dB(monitor mode) |
|
Heart Rate |
Range: 15~350bpm Accuracy: ±1% or ±1bpm, whichever is greater |
|
ECG Gain |
1/4 – 2.5mm/mV@±5%, 1/2 – 5mm/mV@±5% 1 – 10mm/mV@±5%, 2 – 20mm/mV@±5% 4 – 40mm/mV@±5%, Auto |
|
ECG Sweeping Speed |
6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s%±10% |
|
ECG noise level |
<30μ Vp-p |
|
ECG input loop current |
≤0.1μA |
|
Differential input impedance |
≥5MΩ |
|
NIBP |
|
|
Measurement Method |
Intelligent oscillometric method |
|
Parameters |
Systolic, diastolic & mean pressure |
|
Working Modes |
Manual, automatic & STAT (5 min.) |
|
Pneumatic Pressure Range |
0~300mmHg |
|
Cuff Type |
Adult (standard) & child / neonate (optional) |
|
Average Measurement Time |
<90sec. |
|
Inflation Time |
<10sec. (typical adult cuff) |
|
Deflation Time |
<2sec. (typical adult cuff) |
|
Initial cuff inflation pressure |
Adult:<180mmHg Pediatric:<120mmHg Neonate:<90mmHg |
|
Monitoring Range (OPL stands for Overpressure Protection Limit) |
|
|
Adult (OPL 300mmHg) |
Systolic 40~255mmHg Mean 20~215mmHg Diastolic 10~195mmHg |
|
Child (OPL 240mmHg) |
Systolic 40~200mmHg Mean 20~165mmHg Diastolic 10~150mmHg |
|
Neonate (OPL 150mmHg) |
Systolic 40~135mmHg Mean 20~110mmHg Diastolic 10~95mmHg |
|
NIBP Accuracy |
Maximum mean error ±5 mmHg Maximum standard deviation 8 mmHg |
|
|
|
|
|
|
|
Technical Specification MACS 20 |
Sternmed Patient Monitor |
|
RESP |
|
|
Measurement |
Method: Thoracic impedance Range: 0~120rpm |
|
Accuracy |
±2rpm or ±5%, whichever is greater |
|
Alarm Limit Range |
0~120bpm |
|
Alarm Tolerance |
±2rpm or ±5% |
|
TEMP |
|
|
Range |
25.0~45.0℃ |
|
Accuracy |
±0.2℃ |
|
Responding Time |
≤150sec |
|
Unit |
℉ / ℃ |
|
SpO2 |
|
|
Transducer |
Dual-wavelength LED |
|
Wavelength |
Red: 660nm, Infrared: 905nm |
|
Max. Optical Output Power |
Less than 2mW Maximum average |
|
SpO2 Measuring Range |
35~100% |
|
SpO2 measuring accuracy |
±3% (70~100)(RMS of difference) |
|
Low perfusion capability |
0.4~5% |
|
PR |
|
|
Pulse rate measuring range |
30bpm~240bpm |
|
Accuracy of pulse rate |
±2% or ±2bpm, whichever is greater |
|
General Specifications |
|
|
Power supply |
100~250VAC, 50/60Hz |
|
Display mode |
12 inches TFT color LCD |
|
Alarming mode |
Audible & visible alarm |
|
Communication |
Ethernet port |
|
Battery |
Rechargeable Lead-acid battery |
|
Data recording |
Build-in thermal printer (optional) |
|
Classification |
|
|
Safety standard |
IEC 60601-1 |
|
Type of protection against electric shock |
Type BF, CF applied parts |
|
Degree of protection against electric shock |
Class I equipment |
|
Electro-magnetic compatibility |
Group I, Class A |